Welcome to Shenzhen YOURUIAN Technology Co., Ltd.
Sales Hotline 136-8498-6459 admin@youruian.com 中文
CHUNJING Series

Pure Crystal Series Online Gas Purity Analyzer/Dew Point Analyzer

Model: YFZC Key Features: Utilizes a novel gas flow stabilization system and an ultra-stable thermal conductivity sensor, ensuring a long calibration cycle

Sales Hotline
136-8498-6459
💡 Threshold check / alarm setpoint? Quick links to engineer tools: 🧪 Concentration Converter → ⚙️ GB Alarm Calculator →

Product Overview

The Pure Crystal series of online gas purity analyzers primarily operate based on thermal conductivity and other principles, enabling real-time, online monitoring and analysis of multiple gas components. Designed in a standard 4U chassis (customizable), these analyzers incorporate safety barrier technology and are manufactured strictly in accordance with the analyzer’s technical specifications, ensuring safe use in industrial sites. They are novel industrial gas analyzers developed using imported thermal conductivity, electrochemical, ion flow, infrared, ultraviolet, laser, and other technologies combined with new microcontroller technology

Application Scenarios

Product Features Utilizing an ultra-stable thermal conductivity sensor, the calibration cycle is long ●LCD display, Chinese menu, easy to maintain ● Pressure regulating valve, unaffected by fluctuations in inlet pressure ● A drying device can be installed to effectively remove moisture ● Explosion-proof design, explosion-proof grade CT6 ● Can be independently wall-mounted or integrated into a cabinet for sampling rack installation ● Calibration interface, facilitating regular calibration Product Specifications
Detected gas: Combustible gas, toxic gas, oxygen, VOC, dew point; Range selection: 0-1000PPM/99.99%/99.999%/99.9999%/100%, +20 to -80℃
Resolution: 0.001%/0.01%/1PPM, 0.1℃, etc. Detection accuracy: ≤±2%F.S
Repeatability: ≤±1%F.S Stability: ≤±1%F.S/7d
Response time: T90≤30S Sample gas flow rate: 200-2000ml/min
Sample gas pressure: 0.05Mpa≤ inlet pressure ≤0.50Mpa Sample gas temperature: 0-50℃
Ambient temperature: -5℃~+50℃ Environmental humidity: ≤90%RH
Output signal: 4~20mA (customizable upon request, must be specified when ordering) Communication mode: RS232 or RS485 (communication protocol: Modbus RTU)
Operating power supply: 100~240VAC  50/60Hz  1A Outline dimensions: 160mm (width) X 160mm (height) X 280mm (depth)
Opening size: 160mm (width) X 160mm (height) X 280mm (depth) Instrument weight: Approximately 2.5kg
Measurement principle: Contact the manufacturer Measuring medium: Inert gases, H2, CO2, CO, and CnHm, among other common gases (such as those containing SO2, NH3, H2S, etc., must be specified when ordering)
Lifespan: Greater than five years (normal use) Sensor lifespan: Greater than two years (normal use)
Precision pump-aspirated: For use with sample gas pressure at negative or atmospheric pressure (must be specified when ordering)